Summary of problem: Researched and assessed methods for identifying and quantifying vibration, then prototyped and sourced dampening methods within the constraints of the existing device.

Skills used/developed:

  • Problem identification and definition, including creating metrics for quantifying symptoms and evaluating solutions

  • Data acquisition using KEYENCE LS laser scanning system and Starrett materials testing system

  • Rapid prototyping of various dampening methods by hand

  • Compiled and reported results to internal management including engineering management, COO and clinical specialist for approval

​​​

Outcome:

  • Successfully observed “bounce” phenomenon using laser scanning system, the root cause of much of the excess vibration

  • Implemented a small code change in the device controller, mitigated the bounce by improving device actuation timing

  • Recommended a dampening ring which was approved by management and implemented in the next build of clinical product

  • Participated in disposable clinical product build and verification testing in cleanroom environment